Troubleshooting
Continued from page before
Indicator
Possible reason
Solution
By
Continue of:
Compressor is
operating, cryogenic
refrigeration not
sufficient.
Cold head
contaminated.
Remount the cold head.
Pump and flush the cold
head. See chapter
cedure in case of Cryogenic
Refrigerator failure” on
page 6-111
Bruker Service
Cold head defective.
Replace the cold head. See
chapter
of Cryogenic Refrigerator
failure” on page 6-111
Bruker Service
Cryogenic Refrigera-
tor contaminated.
Replace the complete Cryo-
genic Refrigerator. See chap-
ter and
Cryogenic Refrigerator
Parts” on page 4-54
.
Bruker Service