Operating the VIBROCONTROL 1500
VIBROCONTROL 1500
VC1500EN / C103845.002 / V 09 / 14.04.2015
- 49 -
The
Trim
value is calculated according to the following formula:
Sens
nominal
nominal value of the transmission factor 100 mV/g.
Sens
real
actual value of the transmission factor xxx mV/g of the sensor
If the sensitivity is set to
OFF
, the channel is switched off and no
measurement takes place at this channel.
Note:
As long as no sensor is connected to the sensor connection terminals
and the sensitivity is not set to OFF, the instrument will signal a fault.
Fig. 49: Setting for sensor sensitivity
After pushing the
ESC
button to return to the main menu, the query asking
whether the changed values should be saved will appear if any changes
have been made.
Using the buttons
and
select either
No
or
Yes
and confirm by pushing
the button
ENTER
.
Fig. 50: Saving the new settings per measurement channel
Trim
= Sens
nominal
/Sens
real