10
Aesculap
®
TSPACE PEEK/TSPACE
XP
instruments
Preparations at the place of use
►
If applicable, rinse non-visible surfaces preferably
with deionized water, with a disposable syringe for
example.
►
Remove any visible surgical residues to the extent
possible with a damp, lint-free cloth.
►
Transport the dry product in a sealed waste con-
tainer for cleaning and disinfection within 6 hours.
Preparation before cleaning
►
Dismantle the product prior to cleaning, see Disas-
sembling.
►
For mechanical cleaning: Mount the irrigation
connector
4
to the rear end of the rotary nut
3
.
Cleaning/disinfection
Product-specific safety notes on the repro-
cessing procedure
►
Use suitable cleaning/disinfecting agents if the
product is put away in a wet condition. To prevent
foam formation and reduced effectiveness of the
process chemicals: Prior to mechanical cleaning
and disinfection, rinse the product thoroughly with
running water.
►
In case of bone or tissue residues on the product:
Pre-clean the product manually with a brush.
CAUTION
Damage to the product due to
inappropriate cleaning/disin-
fecting agents and/or exces-
sive temperatures!
►
Use cleaning and disinfect-
ing agents according to the
manufacturer’s instruc-
tions which
–
are approved for plastic
material and high-
grade steel,
–
do not attack softeners
(e.g. in silicone).
►
Observe specifications
regarding concentration,
temperature and exposure
time.
►
Do not exceed the maxi-
mum permitted cleaning
temperature of 96 °C.
Summary of Contents for Aesculap TSPACE PEEK
Page 2: ......
Page 3: ...A C D E 1 2 3 4 5 6 7 B...
Page 167: ...165 3 A loosen 6 1 1 D 7 A 2 2 5 3 6...
Page 169: ...167 C A 5 5 D C 6 6 C A 3 loosen 5 2 3 5 2 2 1 a 7 A 7 7 a 2 3 0 8 8 0 A C C a 7 a A a 7 A...
Page 173: ...171 pH 8 www a k i org Rote Brosch re 6 4 3 96 C...
Page 174: ...172 Aesculap TSPACE PEEK TSPACEXP SN304R 20 SN305R SN320R 20 SN304R...
Page 175: ...173 SN305R 20 SN320R 20...
Page 177: ...175 II III 5 IV 5 V...
Page 179: ...177 II III 5 IV 5 V...
Page 181: ...179 FDA CE BBraun Stabimed I 15 1 5 II T C F t I 15 2 pH 9 II 1...
Page 182: ...180 Aesculap TSPACE PEEK TSPACEXP BBraun Stabimed I 35 15 1 5 II T C F t I 15 2 pH 9 II 1...