CAUTION
17
• When the battery charge is low, the actual load may not reach the preset torque limit
value. In this case the auto reverse function will not be activated.
• If a load is continuously applied to the motor handpiece, it may automatically stop to
prevent overheating. In this case, let the motor handpiece cool down.
• If the ambient temperature is low, the alarm may sound when the motor handpiece
rotates, however, this is not a failure. If the alarm does not stop sounding even when
the ambient temperature increases, clean the contra angle head and perform
calibration. (Refer to 6 (2) Calibration)
Reverse rotation
when load continues,
forward rotation
when load is removed
Load lower than
the set torque limit value
Load higher than
the set torque limit value
AUTO REVERSE
The handpiece starts in forward rotation. When a load higher than the torque limit is applied,
the file will rotate in reverse. When the load is removed, the file will return to normal rotation
(forward) automatically.
Reverse rotation
when load continues,
stop when load is removed
Load lower than
the set torque limit value
Load higher than
the set torque limit value
Forward rotation
Returns to forward
Reverse rotation
Forward rotation
Stop
Reverse rotation
Fig. 9
AUTO STOP
The handpiece starts in forward rotation. When a load higher than the torque limit is applied,
the file will rotate in reverse. When the load is removed, the motor rotation stops. If you want it
to rotate again, re-press the ON/OFF Key.
AUTO REVERSE OFF
The motor handpiece stops without reverse rotation. If you want it to rotate (forward-rotate)
again, re-press the ON/OFF Key or re-step.
Fig. 10