8. If you want to stop this process, turn off the power.
(3) Alarm Sound Volume Control
1. Setting alarm sound
You can turn ON/OFF the alarm (load alarm) which sounds near the torque limit value when
the motor handpiece rotates and the alarm during auto reverse (auto reverse sound).
1. Turn off the power
2. While pressing SELECT key, keep pressing POWER key one second or more.
3. The alarm mark and ON or OFF on the LCD panel.
4. Press the SELECT key to select ON or OFF for alarm sound.
5. Display on the liquid crystal panel (LCD) returns to the original state if no operation is
performed for a while.
• This function does not work unless remaining battery capacity is sufficient.
• Perform calibration after cleaning the contra angle head. Residual contamination on
the rotating shaft will impair correct measurement.
• Do not touch or apply a load to the rotating shaft of the contra angle head. It will
obstruct accurate measurement.
• This function cannot completely detect the individual difference of the actual motor
handpiece and contra angle head.
The volume setting is memorized automatically to the last setting and remains
unchanged even if the power is turned off.