24
If the motor handpiece stops due to an abnormality such as a malfunction, overload,
breakage or wrong use, it automatically checks the state of the control unit and detects the
cause of the abnormality and displays an error code on the LCD panel. If an error code is
displayed, turn on the power again and check whether the same error code is displayed. If
the same error code is displayed, take action by referring to the instructions provided in the
“Check/Remedy” column in the following table.
12. Error Code
Contact Brasseler USA
Remove load
Contact Brasseler USA
Leave it as it is until it
cools down.
Contact Brasseler USA
Put the batteries into the
battery chamber, or
replace with new batteries.
Contact Brasseler USA.
Use within the range of
working temperature.
Replace the batteries. If
the heat generates from
the new set of batteries,
malfunction of the circuit
may be suspected.
Contact Brasseler USA.
Replace the motor
handpiece or contra angle
head.
Malfunction of circuit
The motor handpiece is
locked. (at the time of
the auto reverse mode)
Malfunction of circuit
High load was
continuously applied to
the motor handpiece
for a relatively long time.
Malfunction of charger
The voltage of batteries
is too low (“dead”
batteries or not inserted.)
The voltage of batteries
is too high.
(Malfunction of circuit)
Outstanding the range of
working temperature or
break in the thermistor in
the battery section
The batteries generate
abnormal heat.
The operating life of the
motor handpiece or
contra angle has expired
Check/Remedy
Cause
Self-Check
error
Overcurrent
Overvoltage
Overheating of
motor
Charger failure
Low voltage of
batteries
High voltage of
batteries
Outside the
range of
working
temperature
Abnormal heat
generation from
batteries
Beyond the
upper limit
Below the lower
limit
Error
Error code
During
rotation of
the motor
handpiece
At the time
of
Charging
Other
At the time
of
calibration
Summary of Contents for EndoSequence II