Guidance and manufacturer's declaration - electromagnetic immunity
The EndoSequence II is intended for use in the electromagnetic environment specified below. The
customer or the user of the EndoSequence II should assure that it is used in such an environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Immunity test
IEC60601 test level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
±6kV contact
±8kV air
±2kV for power
supply lines
±1kV for
input/output
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 secs
3 A/m
Compliance level
±6kV contact
±8kV air
±2kV for power
supply lines
±1kV line to line
±2kV lines to earth
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
3 A/m
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
EndoSequence II requires
continued operation during power
mains interruptions, it is
recommended that the
EndoSequence II be powered from
an uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.