CAUTION
Contra angle head
F type spray nozzle
SPRAY
20
(1) Lubricating Contra Angle Head
• Lubricate the contra angle head only.
• Apply SPRAY after each use and/or before each calibration and autoclaving
1. Lubricate before each autoclaving or once a
day if you do not autclave, using SPRAY
lubricant. Secure the F type spray nozzle onto
the Spray nozzle.
2. Insert the F type spray nozzle into the rear
part of the contra angle head and lubricate the
head for 2-3 seconds. If the handpiece does
not push into nozzle enough, oil may not go
around into the handpiece and it may flows
backward.
(2) Cleaning Motor Handpiece
When the motor handpiece becomes dirty, wipe it off with a cotton cloth moistened
with rubbing alcohol.
(3) Cleaning Charger
When the charger becomes dirty, wipe it off with a cotton cloth moistened with
isopropyl alcohol.
7. Cleaning
• Do not lubricate the motor handpiece.
• Before mounting the lubricated contra angle head to the motor handpiece, wipe off
excess oil. Stand it on its end or lean it in the proper position for gravity draining.
Mount it after excess oil has been completely drained.
• Hold the contra angle head securely to prevent it from flying off by the pressure
of the spray.
• Hold the spray can upright.
• When cleaning the motor handpiece, do not use any solvent such as benzene
and, thinner.
Fig. 11