Carephone 52+
Programming | en
39
Bosch Security Systems
Operating Manual
F.01U.031.884 V4 | 2009.03
73
Assignment of input
The Carephone 52+ offers the possibility for external inputs.
The function assigned to the reception can be programmed.
The input can be chosen as a closing contact (normally open)
or as an opening contact (normally closed), or controlled by a
resistance. You can program:
X
0 = Ext. activity monitor reset
X
1 = Emergeny call button
X
2 = Service call
X
3 = External alarm
X
9 = Fire
X
B = Motion
Explanation of the motion detection:
With CPC receivers, the alarm type is always set to the alarm
type 10. When an motion detector is installed as an external
receiver, this transmitter will act as an activity monitor reset
when the user is at home, and as a motion detector, when the
user has left. When the user signs out and leaves the house,
the Carephone 52+ activates the motion supervision after 2
minutes. As the user comes back or in the event of a motion
detection, the Carephone 52+ starts a prealarm, that can be
stopped within one minute by pressing the sign in / sign out
button. In the second step, you can determine whether the
closing or opening contact is chosen.
You can program:
X
0 = Closing
X
1 = Opening
X
2 = Impedance is monitored (12k
Ω
when non-operative)
74
Silent alarm
The Carephone 52+ can transmit silent alarms. When this
function is active, then the loudspeaker of the Carephone 52+
is turned off. Only the microphone stays active. The Care-
phone will speak only during programming. You can program:
X
0 = Off
X
1 = On
Table 12.2