Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that is used in such an environment.
Immunity test
IEC 60601 test level
Compliance Level
Electromagnetic environment - guide
Electrostatic
discharge(ESD)
EN 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst,
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
Output lines
± 2 kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5%UT (>95% dip in
UT For 0.5 cycles 40%
UT (60% dip in UT)
for 5 cycles <5%UT
70% UT (30% dip in
UT) for 25 cycles
<5%UT
<5%UT (>95% dip in
UT) for 5 s
<5%UT (>95% dip
in UT For 0.5 cycles
40% UT (60% dip in
UT) for 5 cycles
<5%UT 70% UT
(30% dip in UT) for
25 cycles <5%UT
<5%UT (>95% dip
in UT) for 5 s
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the device requires continued
operation during power mains interruptions,
it is recommended that the device be powered
from an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE:
UT is the a.c. mains voltage prior to application of the test level.