Response of alarm
:
With the file depth, alarm sound from slow to fast, to APEX, the alarm sound becomes constant, over
APEX the alarm sound rapidly changes
Screen display
:
With the file depth, the root canal relative value will change to the apical/APEX appears, Root canal scale
is full, the root canal relative value is “00”; over apical, APEX flashing, Root canal scale is full, the root
canal relative value is “- -”.
4.3 Rotation and settings in different modes
4.3.1 Rotation in different modes
Forward: Rotation is in clockwise motion.
When the load is beyond the set torque limit or reach the apical, it will automatically reverse out,
when the load is removed, it will start normal rotation again.
Reverse: Rotation is in Anti-clockwise Motion.
In the general condition, when the Endo Motor Rotate in counter- clockwise motion will be claim as
reverse motion. The traditional rotary files are drawing from the canal by reverse motion. Therefore,
beep sound will occur when using this function.
Load lower than the set
torque limit value
Reverse rotation when load continues or
reach the apical, stop and return to clockwise
rotation when load is removed.
Load higher than the set
torque limit value,
anti-clockwise
Anti-Clockwise Rotation