14
mode II, in sequence.
Further options for quick start mode:
Machine standby in (Reciprocating/Clockwise Rotation/Anti-Clockwise Rotation) mode
I—Rotation in (Reciprocating/Clockwise Rotation/Anti-Clockwise Rotation) mode I, —standby in
(Reciprocating/Clockwise Rotation/Anti-Clockwise Rotation) mode II, —Rotation in in
(Reciprocating/Clockwise Rotation/Anti-Clockwise Rotation) mode II, in sequence.
Enter the setting:
When the motor is in quick start mode, long press “P” to enter in Quick Start mode setting, to
select OTR
(
reciprocating
)
—Forward (Clockwise Rotation)—Reverse ,adjust by “+/-”key
Adjustable Parameters:
Speed can adjust by “+/-”key
Torque are adjustable in Forward / Reverse mode
The Reciprocating direction and angle are adjustable in OTR mode
(Note: Details please refer to Guide for Quick start function.)
Exit setting:
Long press
“P”
key or short press on/off key to exit the setting.
4.2 Integrated Apex Locator
iRoot pro is equipped with an integrated apex locator.
WARNING:
Combined apical length measurement by using the contra angle and lip clip give only accurate
results when using the original contra angle by manufacturer supplied.
Use only endodontic NiTi files with a metal handle for this type of root canal measurement.
In some clinical cases accurate root canal measurement is impossible! ( for details see chapter
9 “Troubleshooting”)
Electric root canal measurement uses minimal auxiliary current. The values used in the device
are far below the values required in IEC 60601-1. Nevertheless, in rare cases electrical
sensation during the apical length measurement can occur. In this case, do not continue the
treatment with this patient.
4.2.1 Four working mode
There are four types of working mode related to Endo motor an Apex Locator.
Mode, Apex locator control motor how to rotate.