22
Long press “S” key or short press “P” key to exit the setting, and all the settings will be saved
automatically.
CAUTION:
If this function is used, all programs will disappear and return to the originally set values. Record the
present program details if required before you carry out this operation.
4.5.6 Left-hander & Right-hander setting
The program is applicable to Left-hander & Right-hander user.
Long press “S” key to enter into background, short press “S” key until in to Left-hander &
Right-hander option, adjust by “+/-” key for left and right hand use.
Long press “S” key or short press “P” key to exit the setting, and all the settings will be saved
automatically.
4.5.7 Auto Calibration setting
Install the fully lubricated contra angle on the central unit and place it on a flat platform.
When the device is on, Long press
“
S
”
key to enter into other operation settings, short
press
“
S
”
key until in to Auto Calibration. Press
“
+/-
”
key to choose “yes” and enter in
Auto Calibration Mode. The motor starts running from the lowest to the highest speed for
automatic calibration.
During the calibration, the motor rotates with the contra angle. Please don't touch it and wait
for the automatic operation to be completed.
When the calibration is finished, the motor stops rotating and shuts down automatically after
30 seconds.
If you want to stop this operation, press the "S" key for a long time or press the "P" key for a
short time to exit the setting.
CAUTION:
Before the calibration operation, the battery should have sufficient power. It is recommended
that the motor be calibrated after full charge.