15
Mode, Apex locator is working independently, and endo motor is invalid.
Mode, Independent Endo motor, Apex locator is invalid.
Mode, Apex locator and motor work separately.
To short press “S” key to enter the working mode option, to select working mode in the select
box by “+/-”key, when the machine is on.
NOTICE:
In
working mode, motor and apex locator will work at same time, but both work
independently without control to each other. When reach apical, the apex locator only indicate
the apical length, will not control motor to switch to anti-clockwise rotation and reverse out.
In working mode, motor and apex locator will work at same time, apex locator will control
the preparation of motor. Endo motor will auto start, auto reverse out
according to the root canal measurement of the apex locator.
4.2.2 Useful tips for accurate length determination
Gloves and rubber dam are recommended in order to isolate the tooth.
Dry the access cavity with the suction pump or a cotton-pellet.
Avoid any direct contact between the contra angle file and mucosa. Use the silicone sleeve for
the contra angle.
4.2.3 Optional Apical Line
This feature is convenient for users to make marks of the
relative distance from apical point;
The range of markers is from 15 to 27;
According to user labeled apical scale, when reaching marker
position, better visual and sound cues.
Complete following steps to change the marks of apical point
Long press “S” to enter other operation mode, keep short press “S” until it indicates apex
locator working length bar, to adjust by “+/-” key.
NOTICE:
Settings will be saved automatically
This mark for the apical is available in all working mode including QS mode and apex locator
control motor working mode once set.
Set the apical line from 15 to 27