23
Before calibration, be sure to clean the contra angle first, because if there are residual
impurities in the process of use into the contra angle, the calibration will cause deviation.
When calibrating, please don't bring file or any load on the contra angle.
Do not shake the motor during calibration.
Do not do this operation during charging.
5. Cleaning, Disinfection
CAUTION:
No part of the device was disinfected before leaving the factory.
WARNNING
Do not immerse the central unit in the ultrasonic cleaner.
The central unit can only be cleaned with cotton cloth dipped in alcohol.
The metal holder can only be cleaned with cotton cloth dipped in alcohol.
The device disinfectable parts: lip hook, contra angle sleeve, contra angle and file clamp.
The lip hook, contra angle sleeve, contra angle and file clamp have biocompatibility (in line with
EN ISO 10993-1). Before using to every patient, these parts must be disinfected. It is suggested that the
method of high temperature and high pressure steam disinfection be adopted. The parameters of high
temperature and high pressure steam disinfection are recommended as follows:
a) Steam sterilization at 121
°
C (250
°
F) for 20 minutes (placed in a sterilization bag), high
temperature sterilization should not exceed 135
°
C (275
°
F). Steam sterilizer should follow standard
EN 13060.
b) The lip hook, contra angle sleeve, contra angle and file clamp can be repeatedly disinfected.
WARNING:
Except the lip hook, contra angle sleeve, contra angle and file clamp mentioned above, all other
parts for the device can’t be sterilized by high temperature and pressure.
file clamp
lip hook
钩
contra angle sleeve
contra angle