Service Manual Miditron
®
Junior
II
/ ID 1997491 / MJ / 1.0 - June 1998 / Page 18
3. Introduction
3.1.4
Changing Range Remisson Bordes
The boundaries of the reflectance ranges used by
Miditron
®
Junior
II
to compute and output the test results
were derived from rigorous comparative tests carried out by
Boehringer Mannheim with native urine. If required, the
factory-set default ranges for Miditron
®
Junior
II
may be
changed to suit individual laboratories’ requirements.
Results obtained based on individually modified ranges are
flagged with an asterisk * under the "Urinalysis" headline on
the patient report.
The ranges for the parameters pH and SG cannot be
modified; nor can the thresholds for wavelength changes for
pH and ERY.
Operator´s Manual outlines how ranges can be selected.
Boehringer Mannheim can give no guarantee as
to the accuracy of results when range
boundaries have been changed.