Service Manual Miditron
®
Junior
II
/ ID 1997491 / MJ / 1.0 - June 1998 / Page 95
11. Software
"Restore the existing customer
setting"
Customer setting
"Restore the default settings"
Default settings
Once the last file has been sent, the display reads:
Press <Set> to select one of the two display readings:
or
When "Default settings" is selected, all instrument settings
will be reset to factory default settings.
Acknowledge the desired setting by pressing <Enter>.
For structural software updates (change in the version
number before the decimal point) there is no optional
choice. In this case, only the standard setting is available,
as data areas or data structures might have changed.
After an update, the analyzer performs a selftest, deletes
the results and resets the sequence number to 0. Then
calibration is requested, before it changes to the Normal
Mode.
If an error is detected in the course of a software update, an
“error no. xx” (xx=1...12) is displayed. A description of the
error codes is given in Chapter13.
Attention: If the previous software has been deleted, it will
be necessary to load an update. Otherwise the analyzer will
not be operational.