11. Software
11.3
Service/Status Software
Further information see Chapter 7.
11.4
Software update
There are two possibilities to install software updates:
- Load down from the two delivered chip cards
- Loading via service mode with the printer interface
11.4.1
Software update via chip cards
Boehringer Mannheim distributes new versions of software,
and software updates, on two chip cards. To install or
update software, proceed as follows:
Switch off power at the AC power adapter.
Insert Chip Card 1 in the program chip card slot (Fig. 61).
Switch power on again at the AC power adapter. You see
displayed:
Press <Enter> to start the software update, or any other
key to cancel.
The software loaded in the analyzer is cleared:
When the software is loading from Chip Card 1, the status
LED flashes red and you see displayed:
Remove Chip Card 1 and insert Chip Card 2 in the program
chip card slot (Fig. 61) when you see displayed:
The status LED is red.
Press <Enter> when you see the message:
When the software is loading from Chip Card 2, the status
LED flashes red and you see displayed:
You briefly see the message:
Remove Chip Card 2 when you see displayed:
The software version number of the newly loaded software
is displayed briefly.
Fig. 61
"Do you wish to perform a
software update?"
SW update ?
"The old software is being
erased"
erasing Program
"Chip card 1 is being loaded"
loading Program
"Insert chip card 2"
Insert Card 2
"Chip card 2 can be loaded"
Load: <ENTER>
"Chip card 2 is being loaded"
loading Program
"Please Wait"
Please Wait
"Remove chip card 2"
Remove Chipcard
"Version number of the loaded
software"
SW.Vers. x x.xx