G5 Weighing Instrument
10-3
Start-up errors
These error codes can only appear during start-up.
Error
code
Explanation
080
Set up version error
The set-up parameters stored do not match the current program Contact
your supplier.
081
Set-up data error
Indicates faulty set-up. Enter set-up mode, perform the necessary editing
and save the new parameter settings or use the
Set Default Values
function (Maintenance menu).
082
Production parameter error
One or several production parameter is corrupt. Return the instrument for
service. This is a fatal error.
083
Wrong type of Fieldbus Module
The fieldbus module in the instrument doesn’t correspond to the set-up.
Correct the set-up or change module.
084
No Fieldbus Module
There is no fieldbus module installed in the instrument but the set-up
indicates there should be a fieldbus module.
Correct the set-up or install the desired module.
085
Fieldbus Module Error
There is some kind of error when configuring the fieldbus module.
Check the status LED’s on the module, check that the set-up corresponds
to the network configuration, baud rate, addresses etc.