G5 Weighing Instrument
6-15
Clock
These registers are used to read the time and date from the instrument.
Scale Preset tare value
This registers is used to read and write a new preset tare.
Level 1 - 4 value
These registers are used to read and write levels that are supervised by the instrument.
Setpoint 1 - 4 value
The registers are used to read and write setpoints. See description of setpoint function.
Legal settings and status
Describes data of settings for Legal Weighing.
These 10 registers must be read all together in one request.
Reg. no.
Data type:
Explanation
R/W
48600 (1 reg)
Integer
Timestamp on last legal lock change,
year
R
48601 (1 reg)
Integer
Timestamp on last legal lock change,
month (1..12)
R
48602 (1 reg)
Integer
Timestamp on last legal lock change,
day (1..31)
R
48603 (1 reg)
Integer
Timestamp on last legal lock change,
hour (0..23)
R
48604 (1 reg)
Integer
Timestamp on last legal lock change,
minute (0..59)
R
48605 (1 reg)
Integer
Timestamp on last legal lock change,
second (0..59)
R
48606 (2 reg)
Unsigned Integer Lock change reference id
R
48608 (2 reg)
Unsigned Integer Audit Trail counter
R
48610 (1 reg)
Integer
Status bits, Bit 0 Legal Lock (OFF/ON)
0 = OFF, 1 = ON
R
48611 (1 reg)
Integer
Checksum for above 10 registers (using
same algorithm as Modbus CRC-16
checksum)
R
Legal weighing print
Describes a number of registers for a set of Legal Weighing measurement data (on the
last printed weighing).
These 30 registers must be read all together in one request.