G5 Weighing Instrument
6-13
Command register
As this register is read, the answer will always contain only zeros.
There are a number of actions that can be activated in the instrument. The value of this
register (when different from zero) will activate one of these actions, as described in
below.
When an action cannot be performed for some reason (wrong state etc.) an exception
is given as reply. When an exception with code 03 is received the command error
register could be read to get a better error explanation.
Cmd Action activated in
instrument
Description
0
No action
1
Start operation
When the instrument is in ‘Wait for start state’,
this command can be used to start up the
instrument.
2
Enter Remote operation
This command disables keys for the
main weight window on the instrument and start
page/main weight page in web interface. This
means that an external computer is controlling
the instrument or the instrument is controlled
using digital inputs.
3
Exit Remote operation
This command enables the keys and leaves
the remote operation.
4
Disable menu on display
Access to the menu on the instrument is
disabled. Any opened menu window will be
closed and the main weight window will be
shown instead. The INFO-key is disabled, and
the shortcuts to Acc. Weight, Preset Tare and
Levels are no longer present.
N.B. This command has no effect on the web
interface. Note also that this command does NOT
disable the ZERO, TARE, GROSSNET, PRINT,
START and STOP keys.
5
Enable menu on display
Menu via INFO key on the instrument is enabled.
6
Disable menu on Web
interface
Access to the menu via web interface is disabled.
Any opened menu page will be closed and the
start page is shown instead. From the start-page
is it possible to open the Live (main weight) page
but no other pages. The MENU button is disabled
on both start page and Live page. The shortcuts
to Acc. Weight, Preset Tare and Levels are no
longer present on the Live page.
N.B. This command has no effect on the
key/display interface on the instrument. Note also
that this command does NOT disable the ZERO,
TARE, GROSSNET, PRINT, START and STOP
keys