Technical Manual
3-12
Range/Alternatives
Explanation and
<default value>
result of alternatives.
Motion Detect Window
Range:
0 to 999999
Unit:
Measurem. unit
<1 * Resolution>
Motion status is ‘on’ when the weight value is not stable. It goes
off when the weight has been stable for the ‘No motion delay
time’.
Motion condition exists if the weight change during a certain
measuring period is greater than the window ‘Motion Detect
Window’. The measuring period 200 ms. This parameter value
has one decimal more than parameter Resolution, to allow
‘Motion detect w.’ to be smaller than the resolution.
No Motion Delay
Range:
0 to 10.0
Unit: s
<1.0>
Delay in seconds from detection of stable weight until the Motion
status goes off.
Motion Check
Off
On
<Off>
Off: Only zero adjustment is inhibited during motion.
On: Inhibits zero adjustment, taring, and printing during motion.
Min. Printable Weight
Range:
0 to 999999
Unit:
Measurem. unit
<0.0>
Defines the smallest allowed weight value to be printed. 0 gives
no restrictions.
Allow Print Low Weight
Not Allowed
Allowed (with
warning)
<Not Allowed>
Defines if Print command is allowed at Low Weight instrument
error code.
Not Allowed: No Print command will be performed if weight is
less then ‘Min. Weight Print’.
Allowed: If weight is less then ‘Min. Weight Print’ the Print
command will be allowed and a ‘Warning’ will be printed and/or
stored in the Long-term Data Store together with the
printed/stored weight value.
Note: this parameter is only shown if parameter ‘Min. Weight
Print’ is set to more than 0.