G5 Weighing Instrument
9-3
•
fraction of the maximum permissible error
p
ind
From indicator part certificate.
•
maximum value of cable length
per wire cross section
6
wiremax
From indicator part certificate.
Table for Load cell
•
maximum capacity
E
max
From test certificate.
•
minimum dead load
E
min
From test certificate. If not specified, set 0.
•
accuracy class
class
From test certificate.
•
rated output
C
From test certificate or data sheet.
•
maximum number of verification intervals
n
LC
From test certificate.
•
minimum verification scale interval
v
min
or the ratio Y=E
max
/v
min
. (
≥
n
LC
)
Y
From test certificate.
•
minimum dead load output return
DR
or the ratio
Z
=E
max
/(2.
DR
)
Z
From test certificate.
Only for multi interval instruments.
•
input resistance of single load cell
R
LC
From test certificate or data sheet.
•
lower limit of temperature range
T
min
From load cell test certificate.
•
upper limit of temperature range
T
max
From load cell test certificate.
•
fraction of the maximum permissible error
p
LC
From load cell test certificate.
Table for Connecting elements
•
fraction of the maximum permissible error
p
con
Set to 0.5.
The connecting elements must be from the list in WELMEC 2.4. The KIS 3 load
cell mounting kit is found from August 2001, issue 2.