USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 9 of 39
•
Alarm to indicate that at least one pressure sensor is not connected during any process
in which the pump is operating.
•
Overpressure alarm (over 460mmHg), in any of the processes: Preheating, Field
filling, Treatment or Field washing.
o
Acoustic and visual high priority alarm. The pump stops and moves back to
reduce the pressure in the circuit.
•
Pump cover open alarm:
-
Acoustic and visual high priority alarm
-
The peristaltic pump and turbine stop
-
All the clamps will close. Once the pump cover is closed again, the process restarts
where it left off when the cover was opened.