USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 4 of 39
•
Use exclusively with PRS disposable kits. The use of any other disposable kits could
damage the equipment and/or harm the patient.
•
If any of the alarms are constantly activated, stop the treatment, turn off the system
and call Technical Service on +44 1582 834 466.
•
The system should only be used by trained and authorised personnel according to the
instructions in this manual.
•
Portable and mobile RF communication systems may affect the system.
•
WARNING
: The use of accessories and cables other than those specified and supplied
with the PRS system as replaceable parts could lead to an increase in the EMISSIONS
or a decrease in the IMMUNITY of the PRS in terms of Electromagnetic
Compatibility.
•
WARNING:
The PRS system must never be used next to or on top of any other
equipment. Should this be strictly necessary, the system must be monitored by the user
to ensure that it is functioning correctly.
•
Even if there is a fault in one component of the system, or if it is affected by EMC
disturbances, the system will not physically harm the patient. In the worst case, the
treatment will stop.
•
The necessary instructions to maintain basic safety and essential functioning with
respect to electromagnetic disturbances for the expected service lifetime are as
follows: Avoid the use of mobiles around the system.
•
The system's USB port is designed for maintenance, repair and data extraction
purposes, the last being the only action to be carried out by the user.
•
Maintenance and repair operations should be performed by Technical Service
technicians only and under no circumstances shall these activities take place during
operational use of the system.
•
At the end of its life cycle, the system should be managed as electrical waste.
4.2
Precautions
•
Do not tip the system.
•
Danger of explosion. Do not use the system in the presence of flammable gases.
•
Do not submerge the system in water or other cleaning liquids. Wipe the system with
a clean, damp cloth.
•
Do not cover the system with objects that could block its ventilation grilles. This could
cause overheating and functional anomalies in its internal components.
•
The system must be kept in a dry place at a temperature between 10ºC and 35ºC.