USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 29 of 39
•
Alarm for the disconnection or erroneous reading of the bag temperature probe, during
the preheating process.
•
Alarm to indicate that at least one pressure sensor is not connected during any process
in which the pump is operating
-
Overpressure alarm (over 460mmHg), in any of the processes: Preheating, Field
filling, Treatment or Field washing.
o
Acoustic and visual high priority alarm. The pump stops and moves back
to reduce the pressure in the circuit.
•
Pump cover open alarm
-
Acoustic and visual high priority alarm
-
The peristaltic pump and turbine stop
-
All the clamps close. Once the pump cover is closed again, the process restarts
where it left off when the cover was opened.
•
Communication error warning
10
CLEANING AND MAINTENANCE
Precaution
Do not submerge the system in cleaning or sterilisation liquids. The system is not designed
for cleaning by immersion.
Do not clean the system with harsh liquid solvents as these products might damage the casing,
labelling or internal components.
Summary of Contents for COMBAT MEDICAL PRS+
Page 16: ...USER MANUAL BSLAB07_v18_ EN Biosurgical S L part of Combat Medical Group Page 16 of 39...
Page 25: ...USER MANUAL BSLAB07_v18_ EN Biosurgical S L part of Combat Medical Group Page 25 of 39...
Page 26: ...USER MANUAL BSLAB07_v18_ EN Biosurgical S L part of Combat Medical Group Page 26 of 39...