SYSTEM SPECIFICATIONS
ELECTRICAL SPECIFICATIONS
System rating
Model
COMPRS03
COMPRS04
Voltage supply
220-240V ~
100-120V ~
Frequency
50 Hz
50/60 Hz
System peak consumption
1.800 W
1.800 W
2 Timed fuses (6,3 x 32 mm):
2 x 10A
2 x 20A
Power cable
4 m long cable (Type C13) minimum thickness of 1.5 mm
2
(16 AWG)
Interface cables
3 interface cables for temperature probes, 3.6m in length and with a thickness of 3.5mm.
Temperature control
Microprocessor control and electronic temperature sensors
Leakage current
UL 60601-1, IEC 60601-1 and CAN CSA 22.2 certifications
ENVIRONMENTAL CONDITIONS
Environmental conditions for use
Temperature
10ºC -35ºC
(Recommended temperature: 18ºC -25ºC)
Humidity
35 - 75%
Pressure
100-70kPa
Transport and storage conditions
Temperature
–
5ºC - 70ºC
Humidity
0% -90%
Pressure
100-10kPa
Environmental requirements
Compliance with Directive 2011/65/UE(RoHS2)
Compliance with Directive 2012/19/UE (WEE)