USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 11 of 39
4.- If you wish to turn the system off completely, press the MENU key and choose
TURN OFF from the menu. Follow the on-screen instructions for the correct switching
off of the system and, once the screen has gone off, press the off button and the green
light will go off.
7
DISPOSABLE KIT
The system is designed to work exclusively with the disposable kit supplied by Biosurgical
S.L.
The use of other disposable systems not compatible with the system may damage the system
and/or harm the end user.
The disposable kit contains a disposable aluminium heat exchanger, the characteristics of
which are as follows:
•
Priming volume: 200ml.
•
Maximum Pressure: 460mmHg (600mbar)
To learn more about the disposable kit, see the instructions for use of the kit.
Components for the use of the system (pressure transducer, temperature probes, CO
2
Chamber
Clamp, CO
2
Chamber) are included as part of the system.