USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 38 of 39
Table C.8
Manufacturer's guidelines and declaration - electromagnetic immunity - for all devices and systems.
Manufacturer's guidelines and declaration - electromagnetic immunity
.
The PRS
+
system is intended for use in an electromagnetic environment like the one specified below. The client
or user of the PRS
+
system should ensure that it is used in such an environment.
Immunity test
IEC 60601 Standards
test level
Conformity level
Electromagnetic. Guidelines.
Conducted RF
IEC 61000-4-6
Standard
Radiated RF
interference
IEC 61000-4-3
Standard
3V
eff
150KHz to 80MHz
3V/m
80MHz to 2.5GHz
3V
3V/m
80MHz to 2.5GHz
Laboratory
accredited up to
1GHz)
Portable and mobile RF communications devices
should not be used any closer to any other part of
the PRS
+
system, including its cables, than the
recommended working clearance for the
transmission frequency.
Recommended working clearance
d = 1’17
d = 1’17
80MHz to 800MHz
d = 2’33
800MHz to 2.5GHz
Where P is the maximum transmitter output in
watts (W) specified by the transmitter manufacturer
and k is the
recommended working clearance in metres (m).
The field strength of stationary radio transmitters,
based on an on-site electromagnetic study, must be
lower than the compliance level in each frequency
range.
Interference may occur in the vicinity of any
equipment bearing the following symbol:
p
p
p