USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 18 of 39
8.3
Preheating
The preheating screen provides two options:
“Fill Bag”
and
“P
reheating
”
.
Fill Bag
: To start and stop the filling of the bag the "Fill" and "Stop" buttons in the "Fill Bag"
screen must be pressed.
Preheating
: To start and stop the preheating process, the "Start" and "Stop" buttons in the
"Preheating" field must be pressed. Whenever this process is underway, the timer will update
on an ongoing basis. Preheating will take place to the set temperature selected by the user
(43ºC by default), changeable between 20ºC and 45ºC.
The temperature of the fluid in the bag can be checked at any time if the corresponding
temperature probe has been connected. If the probe is not connected to the monitoring cable
and the cable to the system, the preheating process will not be possible.
During the whole preheating process, the temperature control software ensures temperature
stability with an accuracy of ±1ºC with respect to the set temperature.