USER MANUAL
BSLAB07_v18_ EN
Biosurgical S.L., part of Combat Medical Group
Page 28 of 39
The alarm can also be muted for a period of one minute if required by pressing the
corresponding button:
In the case of a malfunction, the system has detection systems and visual and audible alarm
systems to alert system users to hazardous situations. Depending on the situation, the system
has high, medium and low priority alarms.
•
Alarm for over 44ºC and less than 45ºC, for 45 seconds, in the bag probe during the
preheating process or in the irrigation probe during the field filling, treatment and field
washing processes. (working with a SET of 43ºC or less)
-
From 43ºC up to 45ºC, every 0.5ºC there will be a warning message indicating that
"It is not advisable to work with temperatures over 43ºC for this type of treatment"
and confirmation to continue treatment will be required.
-
After 45ºC and up to 48ºC the system will send a message saying: " Ensure that the
temperature in the abdominal cavity is being monitored " and confirmation to
continue treatment will be required.
-
The alarms will adjust with respect to the set temperature as follows:
-
More than 44ºC and less than 45ºC, for 15 seconds, in the irrigation probe, the
alarm will activate one degree over the set temperature selected.
If the temperature selected is lowered again the alarm will stop until the heating
plates start to heat up again and at this point it will set itself at 1ºC above the new
temperature selected.
•
There is an alarm for temperatures over 45ºC in the bag probe during the preheating
process or in the irrigation probe during the field filling, treatment and field washing
processes. (working with a set temperature of 43ºC or less)
•
There is an acoustic and visual high priority alarm if the system does not heat up.
•
If the user increases the set temperature to over 43ºC, the temperature alarm will
activate at 2ºC over the set temperature selected.
•
If the set temperature is lowered again, the alarm will stop until the heating plates start
to heat up again and at this point it will set itself at 2ºC above the new set temperature
selected.
•
There is an alarm for the disconnection or erroneous reading of the irrigation
temperature probe, during the field filling, treatment and field washing processes