TEMPERATURE: The temperature systems can be
tested by these controls. The “HEATER” button enables
the heater and the convection fans. The incubator PV
(Present Value) will be displayed below. The tray cooling
pump can be turned on by pressing the “Tray Pump”
button. The tray temperature will be displayed below
Figure 23: Heater controls
FLUIDICS: The fluid handling manifold is the most
valuable sub-system for automating a Perifusion
experiment. It is composed by the precision peristaltic
pump, and the 8 to 4 manifold. The pump can be tested
by assigning a value to the “Pump Flow-rate” box. The
state of each of the 32 individual valves on the perifusion
can be manually changed by touching the corresponding
LED indicator on the graphical array. An open valve is
represented by an ON indicator. It is not recommended
that you open all valves at once.
Figure 24: Valve controls
At any point, press the BIOREP logo to return to the MAIN MENU
Figure 25: Return to Main Menu