CardioCare2000 Operation Manual
Control panel
▣
Indicator lamp
Indicate the lead connection status.
If LED is illuminated green, connection status is good .
If LED is off, connection status is bad.
In this case, check off which one of lead connection is in lead-off status
through monitor mode output.
Caution: Start printing when LED is illuminated green.
Indicate the battery capacity status: Green and Red colors are used
to display 2 different statuses.
The battery indicator is red when the device is operational without having
an A/C power supply. When an A/C power is supplied, the battery
indicator will be either red or green depending on the charge level. No
color will be shown, if there is no battery installed in the device.
Indicate AC power is operated if LED is illuminated green and not in use if
LED is off.
▣
Short keys
Select signal level(mm/mV) out of 5, 10, 20, aut
(I-aVF:10, V1-V6:5 )
Select printing speed(mm/sec) out of 12.5, 25, 50
Select on or off whether or not to activate the filter that eliminates base
line drift
Select on or off whether or not to activate the filter that eliminates EMG.
ECG signal can be distorted by applying this filter.
This key is excepted to use when network function is added ahead.