CardioCare2000 Operation Manual
4) System Installation
Installation precautions
When installing EKG2000 system, take the following precautions.
-
The equipment should be operated within 10
∼
40
℃
temperature and
30
∼
85
%
humidity.
-
Check the power cord connection and probe should be handled with care.
-
Do not plug several power cords in one electric outlet.
-
Main unit should be installed on an even, level place.
-
Check that the equipment is properly grounded to avoid noise.
-
The entire system settings are stored on the internal memory even when system
power is turned off and on again
-
Avoid excessive shock or vibration that could result in equipment damage.
-
Avoid using in the vicinity of dust and flammable material
.
Power cable connection
The equipment operates when the power cable end connect into the EKG2000 power cord.
Replace the fuse of the equipment if power is not turned on despite that power cord is connected
and battery power is in normal condition. Contact service dept. of Bionet Co.,Ltd if the equipment is
not operated after replacing fuse.
Caution
For the continuous protection against the fire hazard, the fuse must be changed to the
same type and rate of fuse by technician (T3A250V).
For the protection of environment, do NOT throw wasted fuse away in a dangerous place
for the example fire hazard
Summary of Contents for CardioCare 2000
Page 1: ...CardioCare 2000 OPERATION MANUAL Ver 2 41 www ebionet com ...
Page 50: ...50 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 3ch 1rhy ...
Page 51: ...51 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 6ch 1rhy ...
Page 52: ...52 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 12ch rhy ...