35 / 61
Ver. 2.41
CardioCare2000 Operation Manual
Channel form setting
Channel form setting is to adjust the numbers of channel of output form.
This system provides four kinds of channel form of 3ch+1rhy, 6ch+1rhy, 12ch rhy, 60s 1rhy.
As for 3ch+1rhy, 10 seconds of ECG data are recorded in four consecutive segments of 2.5
seconds such as I, II, III for first 2.5 seconds time segment, aVR, aVL, aVF for second 2.5
seconds time segment, V1,V2, V3 for third 2.5 seconds time segment, V4, V5, V6 for last 2.5
seconds segment. And 1 channel rhythm is recorded for 10 seconds in lower part of the output
form.
As for 6ch+1rhy, 10 seconds of ECG data are recorded such as I, II, III, aVR, aVL, aVF for first 5
seconds time segment and V1, V2, V3, V4, V5, V6 for next 5 seconds time segment. And 1
channel rhythm is recorded for 10 seconds in lower part of the output form.
As for 12ch rhy, 12 channel rhythms are simultaneously and continuously recorded in the order
of I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
As for 60s 1rhy, 1 channel rhythm is recorded with 6 lines, 10 seconds to every line, for 60
seconds.
To set the channel form, press FORM key. FORM key toggles between 3ch+1rhy, 6ch+1rhy,
12ch rhy, 60s 1rhy that are displayed on the LCD.
The selected setting is shown on the top line of the output form indicated by
‘3 Ch
1Rhythm Report
’ as for 3ch+1rhy, ‘6Ch 1Rhythm Report’ as for 6ch+1rhy, ‘12Channels
Rhythm Report
’ as for 12ch rhy, ‘1Rhythm Report (60sec) as for 60s 1rhy.
Rhythm channel setting
Rhythm channel setting is to adjust the lead of output form. To set rhythm lead, press LEAD key.
LEAD key toggles between I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 that are displayed on
the LCD. The selected rhythm lead recorded in lower part of the output form for 10 seconds is
the standard lead compared with other channel.
Summary of Contents for CardioCare 2000
Page 1: ...CardioCare 2000 OPERATION MANUAL Ver 2 41 www ebionet com ...
Page 50: ...50 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 3ch 1rhy ...
Page 51: ...51 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 6ch 1rhy ...
Page 52: ...52 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 12ch rhy ...