33 / 61
Ver. 2.41
CardioCare2000 Operation Manual
2) Basic setting
General information
LCD displays current system setting in the order of signal level, printing speed, base line
filter ,EMG filter, heart rate, channel form, rhythm channel when system is turned on . These
settings can be changed by two kinds of methods. First method is to use short keys. System
setting can be easily changed by pressing short keys of control panel that accord with the icons
displayed on the LCD screen. Second method is to use menu. Once, move to menu mode while
LCD displays current system setting by presage
key and then select
“PAT”, “FLT”, “PRN”.
“PAT” can be used to enter the patient information. , “FLT” can be used to select the filter.
“PRN” can be used to adjust printing.
Signal Level Setting
The signal level should be adjusted in case that signal overlaps with neighboring traces or the
amplitude of pulse is too small to interpret.
Signal Level setting is adjustable in four ways. The entire 12 channel is set to each out of
5mm/mV, 10mm/mV, 20mm/mV and aut (Auto gain control) which sets I, II, III, aVR, aVL, aVF to
10mm/mV , V1, V2, V3, V4, V5, V6 to 5mm/mV .
As for 5mm/mV, it indicates that the amplitude of 1mV pulse is amplified into 5mm.
To change the signal level, press SENS
key to select the required one.
SENS
key toggles between 5, 10, 20, aut that are displayed on the LCD.
The selected setting is shown on the bottom line of the output report indicated by
‘all channels:
5mm/mV, all channels: 10mm/mV, all channels: 20mm/mV
’ as for 5, 10, 20 and ‘ I-aVF’:
10mm/mV, V1-V6: 5mm/mV
‘ as for aut.
Display speed setting
Display speed setting is to adjust the width of the output signal. Adjustable value is
12.5mm/s, 25mm/s, 50mm/s. As for 25mm/s, it indicates that ECG signal for 1sec is recorded
in the width of 25mm. So, 12.5mm/s is half the width of 25mm/s, 50mm/s is double the width of
25mm/s. For wide range of signal display, set large value.
To change the printing speed, press SPD
key to select the required one.
SPD
key toggles between 12.5, 25, 50 that are displayed on the LCD.
The selected setting is shown on the bottom line of the output report indicat
ed by ‘12.5mm/s,
25mm/s, 50mm/s’. For your reference, set printing speed to 25mm/s when prints with A4 size.
Summary of Contents for CardioCare 2000
Page 1: ...CardioCare 2000 OPERATION MANUAL Ver 2 41 www ebionet com ...
Page 50: ...50 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 3ch 1rhy ...
Page 51: ...51 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 6ch 1rhy ...
Page 52: ...52 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 12ch rhy ...