46 / 61
Ver. 2.41
CardioCare2000 Operation Manual
3) Monitor mode output
In monitor mode output, the measured ECGs is displayed in real-time.
This output function can be used for the purpose of checking out that the entire channel signal
is active before recording ECGs and monitoring ECGs of the patient for long time.
In this monitor mode, filter setting is fixed as follows ;
BASE LINE
On
Muscle filter
Off
AC Rejection Filter
60hz
Low Pass Filter
40hz
Printing speed can be set to 12.5 mm/s, 25mm/s, 50mm/s and signal level can be set to
5mm/mV, 10mm/mV, 20mm/mV, aut.
Output method
Filter setting value is fixed. Set the required printing speed and signal level and press MON
key. Then, the system starts printing according to the selected speed with the following
message on the LCD.
Press
key or
key to stop printing. Then system stops printing and displays system
setting after indicating following message for 1 second on the LCD.
Output form
In monitor mode output, 12 channel rhythms are printed simultaneously and continuously as
shown below.
Indicated heart rate is the average of previous 8 heart rates.
If the MONI M-CH is set at 3 or 6, please press
key to start printing.
When you press
key again the lead is changed to next group and the printing starts.
If you want to stop in the middle of printing, press the
key.
* MONITOR *
Printing
* MONITOR *
aborted!
Summary of Contents for CardioCare 2000
Page 1: ...CardioCare 2000 OPERATION MANUAL Ver 2 41 www ebionet com ...
Page 50: ...50 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 3ch 1rhy ...
Page 51: ...51 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 6ch 1rhy ...
Page 52: ...52 61 Ver 2 41 CardioCare2000 Operation Manual RECORD REPORT FORM 12ch rhy ...