ExiPrep
TM
16 Pro
Fully Automated Nucleic Acid Extraction System
A4(210X297)
-22-
VI. ExiStation™ Manager Software (continued)
11) After well allocation is complete, press the ‘Store On’ icon (located at the bottom portion of the window)
to turn the Cooling Block and cooling fan on. The Cooling Block maintains the Elution Tube Rack at a
low temperature to keep the eluted nucleic acids and nucleic acid-diagnostic kit mix refrigerated.
12) Click the ‘RUN(
▶
)' button to initiate nucleic acid extraction.
13) After the extraction is complete, remove the samples (Elution tubes and Diagnostic Kit tubes containing
eluted material) and immediately run real-time PCR or keep at -20
℃
~ -80
℃
.
14) Pull out the Base Plate and remove all accessories including Elution tubes, Buffer Cartridges and Tips.
After all accessories are removed, push the Base Plate back completely and close the door.
15) Press the ‘UV-On' button to initiate UV sterilization of the instrument chamber.
16) The UV sterilization process will take 5 minutes. Progress can be verified through the progress bar
displayed on the instrument LCD touchscreen panel.