5
Pure Water Pack
TM
INSTALLATION
l
oCation
1. The
Pure Water Pack
TM
is designed for “under the sink” installation. The system should be mounted vertically, to a secure
wall or cabinet wall, which provides enough room to perform routine maintenance. This includes a minimum of 14-15/16
inches from the chamber end, to allow for removal of the lamp and/or quartz sleeve, a minimum of 6-inches on the backside
of the water purifier, and a minimum of 2-inches free space under the filter housings, to allow for housing removal, when
changing filter cartridges.
Figure 1 -
Bio-Logic
®
Pure Water Pack
TM
- Drinking Water System
2. The water purifier is intended for indoor use only. The water purifier should be protected from the elements and from
temperatures below freezing. The ambient temperature, in the area surrounding the water purifier, should be between
35� F and 100� F.
3. Electrical power supplied to the water purifier
MUST
match power requirements listed on the water purifier. Use of a voltage
surge protector is recommended.
4. The water purifier
MUST
be connected to the cold water line only.
5. The system should be located as close to the faucet as possible. This eliminates the possibility of the purified water being
re-contaminated by bacteria in the water distribution system after the water purifier.
6.
As with any water handling device, the water purifier should be located in an area where any possible
condensation or leakage from the water purifier, any purifier accessory and/or plumbing will not result in damage to the area
surrounding the water purifier.
For added protection, it is recommended that a suitable drain pan be installed under the
purifier. The drain pan must be plumbed to an adequate, free flowing, drain to prevent water damage in the event of
a leak. There are numerous leak detection/flood stop devices, available on the market today, designed to stop the flow
of water, reducing the chance of water damage due to leakage. For more details regarding leak prevention and/or
limiting damages due to leaks, please contact the factory.
7.
The system should be located within 5-feet of an electrical receptacle. The receptacle must be a properly
installed, grounding type, receptacle that has Ground Fault Circuit Interrupter (GFCI) protection.