background image

12

Q

uartz

 s

leeVe

 i

nstallation

, C

leaning

 

or

 r

ePlaCement

Figure 8 - Quartz Sleeve Cleaning or Replacement

Step 6 - Remove Quartz Sleeve

Step 5 - Remove Washer & O-Ring  

from each end

Step 4 - Remove Gland Nut From Each End

IN ORDER TO PERFORM THIS TASK, BE SURE TO WEAR THE FOLLOWING SAFETY 

EQUIPMENT: SAFETY GOGGLES OR A FACE SHIELD, AS WELL AS GLOVES

1. 

  

  Disconnect power to water purifier.

•  For 120v 50/60Hz version using ballast no. 10-0518A or 230v 50/60hz version using ballast no. 10-0514A, remove 

the ground connector from the chamber, and go to the next step.

2.  Shut off water supply to water purifier via inlet and outlet shut off valves and drain chamber, if possible.
3.  Follow the steps in 

“Lamp Installation or Replacement”

 to remove lamp. 

Lamp and quartz sleeve are easily damaged. Exercise care when handling.

4.  Unscrew static gland nuts from each end of the chamber. Avoid striking quartz sleeve with static gland nut.
5.  Remove Teflon

®

 washer and O-Ring from both ends of quartz sleeve. Teflon

®

 washer will sometimes remain within the static 

gland nut. If so, remove Teflon

®

 washer from static gland nut before proceeding.

6.  Carefully remove quartz sleeve from chamber. 

NOTE: It is advisable to support the quartz sleeve on the opposite end 

with your finger so that it does not drop to the bottom of the chamber as it slides into the chamber.

7.  Once the quartz sleeve is removed, clean with alcohol or a mild, non-abrasive detergent. Stubborn stains usually can be 

removed with a dilute hydrochloric acid. 

NOTE:

 

Follow all manufacturer’s instructions and precautions 

when handling chemicals.

8.  Reassemble in reverse order. Make sure the quartz sleeve protrudes an equal distance past each threaded nipple. Be sure 

O-Rings are placed on quartz sleeve before Teflon

®

 washer.

9.  Tighten static gland nuts firmly by hand only, 

DO NOT USE HAND TOOLS

. Tightening with hand tools is likely to cause 

quartz sleeve to break.

10. 

Slowly

 restore water supply to water purifier and check for leaks.  

(Continued on Page 13)

5.  Reinstall lamp in reverse order.

  Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to water purifier 

until lamp and both 

EASY-OFF

TM

 End Caps have been properly reinstalled.

6.  Reattach the ground connector to the grounding tab of the chamber, and restore power to the system.

Summary of Contents for PURE WATER PACK BIO 1.5 PWP

Page 1: ...SA Models Ballast 120v 50 60 Hz Ballast 230v 50 60Hz BIO 1 5 PWP 10 0518A 10 0514A Extensive Product Information Available at ultraviolet com buyultraviolet com Installation Operation Maintenance Read...

Page 2: ...ISTRATION 20 WARRANTY 20 Patent Notice 20 These instructions generally describe the installation operation and maintenance of the Bio Logic Pure Water PackTM Questions that are not specifically answer...

Page 3: ...to determine and validate the suitability of this equipment for use in the user s system or process No warranty or representation is made by the manufacturer with respect to suitability or performanc...

Page 4: ...osing it to the germicidal lamp Exposure time is related to the flow rate the higher the flow rate the lower the exposure time or the lower the flow rate the higher the exposure time The ultraviolet i...

Page 5: ...ter purifier MUST be connected to the cold water line only 5 The system should be located as close to the faucet as possible This eliminates the possibility of the purified water being re contaminated...

Page 6: ...til it is firmly seated The valve is CLOSED in this position 4 Open the cold water supply valve or main valve check for any signs of leaks around valve connection 5 Install John GuestTM Tubing Connect...

Page 7: ...r b Run the precut 24 tubing from the outlet of the carbon filter to the inlet of the Bio Logic water purifier c Run the tubing from the faucet to the outlet of the Bio Logic water purifier Measure cu...

Page 8: ...ter to run for 10 minutes to flush the filter and system Figure 4 Bio Logic Pure Water PackTM Install 9 Plug ballast into an approved Ground Fault Circuit Interrupt GFCI receptacle Where a 2 prong or...

Page 9: ...let Corporation maintains a qualified team of application specialists that can provide any technical assistance which maybe required during the use of this system For any questions concerns or technic...

Page 10: ...dal lamp operation This provides an indication of lamp operation and does not indicate the level of ultraviolet intensity or transmission through the water 2 To ensure proper operation of the Bio Logi...

Page 11: ...ir bubbles Turn off the faucet and check between the filter housing and head for leaks 12 It is recommended when installing a new carbon filter to allow the water to run for 10 minutes to flush the fi...

Page 12: ...so remove Teflon washer from static gland nut before proceeding 6 Carefully remove quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end with your finger so...

Page 13: ...Carefully discard all pieces of the broken quartz sleeve 8 Inspect the large O Ring used to seal the chamber and the chamber head Make sure the O Ring is seated properly between the chamber head ring...

Page 14: ...power Verify that the purifier is connected to a live power source Water leaking from purifier Cracked or broken quartz sleeve Shut down purifier drain and replace quartz sleeve See Quartz Sleeve Ins...

Page 15: ...Rate GPM 1 5 Inlet Outlet Size 1 4 Tubing to 3 8 MNPT John GuestTM Connector Number of Lamps 1 Lamp Model No 05 1366 R Quartz Sleeve Yes Length 14 1 2 Width 5 3 4 Height 18 Shipping Weight 18 Lbs Vol...

Page 16: ...16 DIMENSIONAL DATA Figure 10 Dimensional Data All specifications dimensional data etc are approximate and subject to change without notice...

Page 17: ...TM Ballast 12v DC 1 10 1201A 3 STER L RAY Lamp 1 05 1366 R 4 Static Gland Nut 2 25 1492D 5 Teflon Washer 2 25 3023A 6 Rubber O Ring Quartz Sleeve 2 00 1023A 7 CRYSTAL CLEARTM Quartz Sleeve 1 15 1700A...

Page 18: ...7 6 Screw 10 16 x Long 8 50 0096 7 Filter Cartridge 5 Micron Pleated 1 25 4508 8 Filter Cartridge Carbon 1 25 4509 9 Male Connector x3 8 NPTF John GuestTM 4 25 4513 10 Pipe Nipple 3 8 x1 Long PVC 1 25...

Page 19: ...19 MAINTENANCE NOTES...

Page 20: ...attempt has been made to determine the patent status of applications illustrated or described in this publication Inclusion in this publication of any design or method of use which may be patented is...

Reviews: