14
13. Remount the water purifier to system-mounting bracket, by firmly pressing the purifier into the chamber brackets.
14. Reinstall the tubing; see “
Tubing Installation
” for the proper method of installing the tubing. Be sure that the tubing from
the outlet of the carbon filter goes to the inlet of the water purifier, and the tubing from system faucet goes to the outlet of the
water purifier.
15. Lift the handle, of the system faucet, and lock the faucet in the “ON” position. Slowly open the supply valve, and allow the
system to fill with water.
16. When water begins to flow from the faucet, allow water to run for several minutes until it is clear with no air bubbles. Turn
off the faucet and check the water purifier for leaks.
Germicidal ultraviolet rays are harmful to eyes and skin. Do not restore power to water purifier until
lamp and both
EASY-OFF
TM
End Caps have been properly reinstalled.
17. If no leaks occur, reinstall lamp by following the steps in “
Lamp Installation or Replacement
” section.
d
isPosal
of
m
erCury
a
dded
l
amPs
Germicidal ultraviolet lamps, like standard fluorescent lamps contain small amounts of mercury. Mercury added
lamps should not be placed in the trash. Dispose of properly. For further information regarding the disposal and
recycling of lamps containing mercury, along with Federal and State requirements visit www.lamprecycle.org. For
more information on germicidal ultraviolet lamps, visit Ultraviolet.com or BuyUltraviolet.com.
TROUBLESHOOTING
Always disconnect power to the water purifier before performing any service or maintenance.
IMPORTANT: This unit is to be serviced ONLY by qualified, and appropriately licensed, personnel.
Table 2 - Troubleshooting
Problem
Possible Cause
Corrective Action
Purifier not operating
No electrical power
Verify that the purifier is connected to a live power source.
Water leaking from purifier
Cracked or broken quartz sleeve
Shut down purifier, drain, and replace quartz sleeve. See
“Quartz Sleeve Installation, Cleaning or Replacement”
or
“Replacement of Broken Quartz Sleeve”
in the
“Maintenance”
section.
Quartz sleeve sealing O-Ring (s) worn,
damaged
Shut down purifier, drain, and remove static gland nut, replace
sealing O-Ring. See
“Quartz Sleeve Installation, Cleaning or
Replacement”
in the
“Maintenance”
section.
Poor, or loose, connections or fittings
Tighten suspect connection or fitting; or shut down purifier, drain,
and remove fitting or connection. Clean threads; reapply thread
sealing tape and reinstall.
Water leaking from fitting
Tubing may not be installed properly
See “Tubing Installation” for the proper method for installing
tubing.
Water cloudy and/or has air bubbles
Air in system
Open system faucet, allow water to run, until water clears and has
no air bubbles.
Water contains black particles
New carbon filter cartridges contain
activated carbon, in a powder from
Open system faucet; allow water to run, purging the carbon
particles from the cartridge, until water is clear.
Water pressure dropped
Filters installed for longer than 12
months
Replace both filter cartridges. See “
Filter Cartridge
Replacement.
”
Water quality poor, filters clogged
Poor purifier disinfection performance
Quartz sleeve fouled
Clean quartz sleeve, see
“Quartz Sleeve Installation, Cleaning
or Replacement”
in the “
Maintenance
” Section.
Germicidal lamp output depreciating
Replace lamp, as it nears its end of life
(EOL)
. See “
Lamp
Installation or Replacement
” in the “
Maintenance
” section.
Germicidal lamp not functioning
Replace lamp. See “
Lamp Installation or Replacement
” in the
“
Maintenance
” section.
Low input voltage
Verify input voltage to purifier.
Change in water quality
Have water tested to confirm that it does not exceed maximum
recommended concentration levels for use with this purifier.
Summary of Contents for PURE WATER PACK BIO 1.5 PWP
Page 19: ...19 MAINTENANCE NOTES...