Calibrating and adjusting the temperature and humidity control
11.5.3 Adjusting the humidity controller
1.
Enter the menu “Inp.Val. correct.“
(see steps 1 to 0, chap. 11.3.1).
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service Level
Inp.Val. correct.
Analogue Inputs
Analogue Outputs
Bin. Inputs
Bin. Outputs
Anal. Inp. mTRON
Anal. Out. mTRON
Bin. Inp. mTRON
Bin. Out. mTRON
Memory Display
↓
Use the buttons
or
to select the menu item
“Analogue Input 4” and confirm with <ENTER>.
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service Level
Inp.Val. correct.
Analogue input 1
Analogue input 2
Analogue input 3
Analogue input 4
Analogue input 5
Safety controller
↓
Use the buttons
or
to select a menu item and confirm
with <ENTER>.
For each menu item a submenu will be opened.
Use the buttons
,
,
and
to key in the values de-
scribed below and confirm with <ENTER>.
State: On
Start value X: Value of “Analogue Inputs“ / “4.” at the
lower set humidity (e.g. +40.0) (step 0, chap.
11.5.2).
Start value W: Value of reference measuring instru-
ment at the lower set humidity (e.g. 42.2) (step 0,
chap. 11.5.2).
End value X: Value of “Analogue Inputs“ / “4.” at the
higher set humidity (e.g. +65.3) (step 0, chap.
11.5.2).
End value W: Value of reference measuring instrument
at the higher set humidity (e.g. 67.5) (step 0, chap.
11.5.2).
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service Level
Inp.Val. correct.
Analogue input 1
Analogue input 2
Analogue input 3
Analogue input 4
Analogue input 5
Safety controller
Analogue input 4
State
Off
Start Value X
+
40.0
Start Value W
+
42.2
End Value X
+
65.3
End Value W
+
67.5
↓
Push <EXIT> four times to return to the basic display.
The corrected value appears on the controller display
after 5 seconds.
Compare the displayed humidity values of the
controller display and the reference measuring
instrument. The deviation must be less than the
confidence criteria (chap. 11.5.2).
W
X
Service Manual KBF (E5.2) 04-2015