Repair work at electric and electronic components
With the options
Analog outputs for temperature and humidity
Alarm outputs for temperature and humidity
Two-channel pen recorder
following replacement of the controller display (-3N1) you need to record again the
mTRON setup. When ordering a spare part, therefore always indicate the type and the
corresponding serial no. of the climatic chamber. You will then receive a complete set
with the mTRON modules corresponding to the required data. Send the removed mTRON
modules back to BINDER-Service. Prior to sending, inscribe on the modules the date
and, if the order was issued by BINDER GmbH Tuttlingen, also the “SAB no.”.
8.10 Replacing individual mTRON relay modules
Spare parts:
mTRON relay module 4015 (-11N1), (order no. 5014-0075)
mTRON Analog output module 4025 (-12N1), (order no. 5014-0074)
Necessary
tools:
Crosstip screwdriver, size 2
Small slotted screwdriver
Replacing individual mTRON modules:
The replacement of individual mTRON modules is made easy by the “Plug & Play” func-
tion. In each module is a storage board (C) containing the configuration data for the LON-
bus addressing. When replacing a module, you only need to place the storage board onto
the new module, and it then works with the original configuration data.
CAUTION!
If you don’t replace the storage board when replacing a module, the module must receive the appropri-
ate data set, according to the optional equipment of the unit.
Only replace a module with one of the same type (e.g.
relay module 4015
with
relay module 4015
)
If you place the storage board at a module with a different functionality, the network will not recognize
the individual options. And in this case the data on the storage board will be deleted!
86
Service Manual KBF (E5.2) 04-2015