Repair work at the humidifying and dehumidifying system
7. Open the blue fastener of the holding band
(steam boiler fixation).
8. The insulation of the steam cylinder can now be pushed
over the lid, and the cables.
Make sure that the cable of the capillary tube sensor is
carefully passed through the opening.
9. Remove the screws on the lid of the steam boiler.
The lid of the steam boiler can now be removed.