Repair work at the outer door
T (°C)
0
1
2
3
4
5
6
7
8
9
10
-10
96.086 96.478 96.870 97.262 97.653 98.045 98.436 98.827 99.218 99.609 100.000
0
100.000 100.391 100.781 101.172 101.562 101.953 102.343 102.733 103.123 103.513 103.902
10
103.902 104.292 104.681 105.071 105.460 105.849 106.238 106.627 107.016 107.404 107.793
20
107.793 108.181 108.570 108.958 109.346 109.734 110.122 110.509 110.897 111.284 111.672
30
111.672 112.059 112.446 112.833 113.220 113.607 113.994 114.380 114.767 115.153 115.539
40
115.539 115.925 116.311 116.697 117.083 117.469 117.854 118.240 118.625 119.010 119.395
50
119.395 119.780 120.165 120.550 120.934 121.319 121.703 122.087 122.471 122.855 123.239
60
123.239 123.623 124.007 124.390 124.774 125.157 125.540 125.923 126.306 126.689 127.072
70
127.072 127.454 127.837 128.219 128.602 128.984 129.366 129.748 130.130 130.511 130.893
80
130.893 131.274 131.656 132.037 132.418 132.799 133.180 133.561 133.941 134.322 134.702
90
134.702 135.083 135.463 135.843 136.223 136.603 136.982 137.362 137.741 138.121 138.500
100
138.500 138.879 139.258 139.637 140.016 140.395 140.773 141.152 141.530 141.908 142.286
Table 1: Temperature – resistances table of temperature sensor Pt100
A
C
F
E
B
D
Figure 11: Checking and replacing the interior side of the outer door
A
Electric mounting plate
B
Interior side of the outer door
C
Door hinge with cover
D
Grounding cable
E
Wires of triangle instrument box
F
Cable for door heat conductor and temperature sensor
Service Manual KBF (E5.2) 04-2015