Calibrating and adjusting the temperature and humidity control
11.4 Setting the door heating correction
1.
Enter the menu “Inp.Val. correct.“
(see steps 1 to 0, chap. 11.3.1).
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service Level
Inp.Val. correct.
Analogue Inputs
Analogue Outputs
Bin. Inputs
Bin. Outputs
Anal. Inp. mTRON
Anal. Out. mTRON
Bin. Inp. mTRON
Bin. Out. mTRON
Memory Display
↓
Use the buttons
or
to select the menu item
„ Analogue Input 2” and confirm with <ENTER>.
Konfiguration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service-Ebene
Meßwertkorrektur
Eingang Kanal 1
Eingang Kanal 2
Eingang Kanal 3
Eingang Kanal 4
Eingang Kanal 5
Überwachungsregler
08:43:55 15.05.00
↓
Use the buttons
or
to select a menu item and confirm
with <Enter>. For each menu item a submenu will be
opened.
Use the buttons
,
,
and
to key in the values
described below and confirm with <ENTER>.
State: On
Start value X: 10.0
Start value W: 8.5
End value X: 90.0.
End value W: 88.5
Konfiguration 2
Analogue inputs
Analogue outputs
Limit comper ators
Donor - functions
Contro ller
Service Level
Service - Ebene
Meßwertkorrektur
Analogue input 1
Analogue input 2
Analogue input 3
Analogue input 4
Analogue input 5
Safety controller
Eingang Kanal 2
Status
Ein
Anfangswert Ist
+
10.0
Anfangswert Soll
+
8.5
Endwert Ist
+
90.0
Endwert Soll
+
88.5
08:43:55 15.05.00
↓
Push <EXIT> four times to return to the basic display.
W
X
Service Manual KBF (E5.2) 04-2015
129