Calibrating and adjusting the temperature and humidity control
11.3.1 Resetting the temperature controller adjustment
Before the unit can be re-adjusted the current adjustment has to be reset and the input value correction has
to be switched off.
1.
Starting from Normal Display push <CONFIG>.
W
X
↓
Use the buttons
or
to select the menu
“Configuration 2” and confirm with <ENTER>.
User-Settings
Choose variation
Parameters
Configuration 1
Configuration 2
↓
Use the buttons
,
,
and
to key in the
System-Code “+02802” and confirm with <ENTER>.
System - Code ?
+02802
↓
“
Use the buttons
or
to select the menu item
“Service Level” and confirm with <ENTER>.
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
↓
122
Service Manual KBF (E5.2) 04-2015