Repair work at the humidifying and dehumidifying system
9.1.3 Freshwater supply
CAUTION
Calcification of the Humidification module.
Damage to the unit.
Operate the unit with deionized (demineralized) water
only
.
•
Supply pressure 1 to 10 bar when connecting to a water pipe
•
Type of water: deionized (demineralized) water.
•
Water intake temperature NOT below +5 °C / 41°F and not exceeding 40 °C / 104°F.
•
The water intake should be provided with a shut-off slide or water-tap.
•
Protect the water supply at one side with the delivered hose clamp.
Types of suitable water quality
•
Deionized water from a water treatment installation already existing at the customer's site.
•
Water treated by the optional water treatment system BINDER
Pure Aqua Service
.
BINDER GmbH is not responsible for the water quality at the customer.
Any problems and malfunctions that might arise following use of water of deviating
quality are excluded from liability by BINDER GmbH.
We recommend using the optional water treatment system BINDER
Pure Aqua Service
.
clean deionized water
Figure 47: Water samples out of the steam boiler
polluted water
96
Service Manual KBF (E5.2) 04-2015