Problem: Unit does not turn on.
Cause
Action
Insufficient power supply when
using the power adapter.
Check the plug connection; if necessary, try
the mains connection at other sockets.
Flat batteries
Replace the batteries.
Incorrectly inserted or missing
battery.
Verify that the batteries are inserted correctly
(see "Inserting and replacing the batteries",
page 22).
Test strip is inserted with the
wrong side or not completely.
Insert the test strip, contacts first, into the slot
on the unit. Ensure that the front side of the
test strip is facing you (see "Test strips", page
19).
Unit is defective
Contact customer service.
Problem: After inserting the test strip into the unit and applying the blood,
the test does not start.
Cause
Action
Insufficient quantity of blood
Repeat the test using a new test strip and lar-
ger blood drop.
Test strip is defective
Repeat the test using a new test strip.
Blood was applied when the unit
was switched off.
Repeat the test and apply the blood when is
flashing.
The unit's basic settings were
changed and the change was
not completed (see "Basic
settings", page 23).
Press the "SET" button until "OFF" is dis-
played.
Repeat test.
Unit is defective
Contact customer service.