Beurer BGL60 mg/dL
49
No.
Cause
Action
Err 06 Deflation rate is too low.
Repeat the measurement. Contact our
customer service department if the
problem persists.
Err 07 Systolic pressure exceeds the
measurement range.
Repeat measurement without
moving and without speaking.
Err 08 Cuff pressure exceeds
the measurement range.
Repeat measurement without
moving and without speaking.
Err 09 Cuff pressure is
insufficient, even with
maximum pump pressure.
The blood pressure is higher than the
measurement range.
Repeat measurement. Consult a
physician if the problem persists.
Err 17 No pulse detected.
Repeat measurement. Consult a
physician if the problem persists.
Err 31 Internal communication error.
Switch off the unit and switch it on
again.
Err 32 Error from moving too much dur-
ing measurement.
Do not move or speak during the
measurement.
Err 41 Pulse amplitude changes sud-
denly.
Repeat measurement without
moving and without speaking.
Err 58 Internal communication error.
Switch off the unit and switch it on
again.
Err … Every error message not noted
in this list is an internal commu-
nication error.
Take the batteries out of the unit for a
few seconds and then put them back.